X-Ray Imaging

Structural & Morphology Characterization (X-ray analysis)

NanoCT (Xradia Ultra 810 X-ray microscope) enables the non-destructive structural characterization of three-dimensional samples with resolution down to 50 nm. In this non-invasive X-ray microscopy technique, the sample is positioned in the focal spot of the X-ray beam and 2D X-ray radiographs of the sample are acquired at different angles. The transmitted X-rays are magnified by a Fresnel zone plate objective lens before reaching the X-ray camera (scintillator optically coupled to a CCD detector). For low absorption samples, a Zernike phase ring inserted into the beam path produce an edge enhancing phase contrast effect. Thus, the nanoCT can be adjusted for absorption and phase contrast and can be operated with a large field of view (FOV: 65 μm, resolution: 150 nm) or at high resolution (FOV: 16 μm, resolution: 50 nm). The 2D projections are then reconstructed which results in a visualization of the sample in 3D. Through further processing, the data can be segmented and quantified. The in situ nanomechanical testing device (with high mechanical stability and a maximum force of 0.8 N), allows to investigate the inner structure of materials under load. High quality nanotomography is very much dependent on the sample. Ideally, the sample is cylindrical, fits the field of view and its thickness is less than twice the X-ray absorption length of the sample material.

i
@
          provided at NFFA-Europe laboratories by:
KIT
Germany
@
          provided at Large Scale Facilities by:
SOLEIL
France
KIT
Germany
NanoCT Xradia 810 Ultra
The nanoCT Xradia 810 Ultra enables the non-destructive structural characterization of three-dimensional samples with resolution down to 50 nm. In this non-invasive X-ray microscopy technique, the sample is positioned in the focal spot of the X-ray beam and 2D X-ray radiographs of the sample are acquired at different angles. The transmitted X-rays are magnified by a Fresnel zone plate objective lens before reaching the X-ray camera (scintillator optically coupled to a CCD detector). For low absorption samples, a Zernike phase ring inserted into the beam path produce an edge enhancing phase contrast effect. Thus, the nanoCT can be adjusted for absorption and phase contrast and can be operated with a large field of view (FOV: 65 μm, resolution: 150 nm) or at high resolution (FOV: 16 μm, resolution: 50 nm). The 2D projections are then reconstructed which results in a visualization of the sample in 3D. Through further processing, the data can be segmented and quantified. The in situ nanomechanical testing device (with high mechanical stability and a maximum force of 0.8 N), allows to investigate the inner structure of materials under load. High quality nanotomography is very much dependent on the sample. Ideally, the sample is cylindrical, fits the field of view and its thickness is less than twice the X-ray absorption length of the sample material.
Cr-anode
5.4 kev
Fields of view: 1) Large field of view: 65 um x 65 um, maximum resolution 150 nm 2) High resolution: 16 um x 16 um, maximum resolution 50 nm
Large field of view: 150 nm High resolution: 50 nm
180 degrees; 140 degrees for mechanical in situ scan
- Cylinder-shaped are preferred but not mandatory. - Thickness should be less than twice the X-ray absorption length of its material (X-ray absorption length can be calculated here). - Sample should fit the field of view (65 μm or 16 μm).
in situ nanomechanical testing: Maximum force: 0.8 N Load modes: compression, indentation, tension
.tiff, .txrm